Sodium analyzer 7 Menu descriptions
OI/ASO550–EN Rev. G 47
Menu Comment Defaults
Solution Detection Enables / disables the solution detection function.
Flow Meter Note. Single-stream wet-sections only.
Enables / disables flow measurement. Disabled
Flow Meter 1 – 3 Note. Multi-stream wet-sections only.
Enables / disables flow measurement for the selected stream. Disabled
Stream Settings Note. Multi-stream wet-sections only.
Sample Time Sets the how long each stream is sampled. 15 min.
Sequence Sets the order in which the streams are sampled, a maximum of 8
entries are permitted. A 0 indicates the end of the sequence.
An example sequence for streams 1 to 3 is: 1,2,3,1,2,3,0 where the 3
streams are sampled consecutively.
1,2,3,1,2,3,1,2,1
(3-stream)
Stream 1 (3) Enables / Disables the selected stream (disabled streams are not sampled).
Man. Valve Control The solenoid valve can be energized manually. The current
concentration value is shown for reference. When the Man. Valve
Control parameter is exited, the valve returns to the closed state.
Note: Press the key to open the valve manually, press the key
to close the valve manually.
Low Standard Valve Enables the Low Standard Valve for manual control.
High Standard Valve Enables the High Standard Valve for manual control.
Stream Valve Note. Single-stream wet-sections only.
Enables the Stream Valve for manual control.
Stream Valve 1 – 3 Note. Multi-stream wet-sections only.
Enables the selected Stream Valve for manual control.
Regeneration Valve If fitted, enables the Regeneration Valve for manual control.
Reset to Defaults Resets all wet-section parameters to their default values. Disabled