Navigator 500
Sodium analyzer 7 Menu descriptions
46 OI/ASO550–EN Rev. G
7.1.2 Sensor Setup
Used to set the wet-section tag, measurement units, operational and stream ranges and filter type
and enable / disable the Flow Meter (optional) and Manual Valve Control functions.
Menu
Sensor Setup
Exit
Select
Menu Comment Default
Sodium 1 – 4 The identity of the sodium wet-section being set up (if a sodium
multi-steam wet-section, Sodium only is displayed).
Tag Note: Single-stream wet-sections only.
Sets the (colored) 16-character alphanumeric wet-section tag
displayed on Operator Pages.
TAG 1
Stream 1 (3) Tag Note: Multi-stream wet-sections only.
Sets the 16-character alphanumeric tag for each selected stream
displayed on Operator Pages.
Units Selects the measurement units:
ppb / µg/l / µg/kg. ppb
Range High Note. Single-stream wet-sections only.
Sets the range span in Chart and Bargraph views.
Stream 1 (3)
Range High
Note. Multi-stream wet-sections only.
Sets the range span in Chart and Bargraph views.
Range Low Note. Single-stream wet-sections.
Sets the range zero in Chart and Bargraph views.
Stream 1 (3)
Range Low
Note. Multi-stream wet-sections only.
Sets the range zero in Chart and Bargraph views.
Filter Type Selects the filter type:
Off / Min / Max / Average / Rolling Point.Off
Filter Time Sets the filter (input) time (5 to 100 seconds). 5 seconds
Wetting Routine Enables / disables the wetting routine. The wetting routine is initiated
when a No Samples situation is detected. It energizes the high
calibration solution valve for 1 minute every 24 hours.
The wetting routine is terminated if sample returns.
Continued on next page…