Navigator 500
Sodium analyzer 7 Menu descriptions
OI/ASO550–EN Rev. G 55
7.1.7 Media Card
Used to set the card on / off status, select the process data to be logged, enter file configuration
selection and save details and to format the media card.
Media Card level menus are enabled only if an optional media card module is fitted.
Menu
Media Card
Exit
Select
Menu Comment Default
Card Status Selects card status, Online / Offline.Online
Process Data
Channel 1 to 6 Selects the source to be logged – refer to Section 7.2, page 58 for
sources.
Sampling Time Selects the sampling duration time:
5 / 10 / 30 seconds
1 / 5 / 10 / 30 minutes
1 hour
5 s
Data Logging Enabled / Disabled data logging.
In Enable mode data can be written to internal / external media.
In Disable mode, data is prevented from being written to internal /
external media.
Enabled
Save Configuration
Select File Name (8) Enables a user-selected filename to be specified for the current
configuration to be saved.
Load Configuration
Select File (8) Selects the configuration file to be loaded from a list of previously
saved files.
Quick Format Disabled / Enabled.
Runs a quick format routine on the SD card / USB stick inserted into
the transmitter’s card reader.
Disabled

Table of Contents