Navigator 600 Series Analyzers
Silica and Phosphate Appendix B – Acyclic Parameter Mapping
IM/NAV6PBS Issue 1 15
Stream 1 tag 1 61 Simple OctetString 20 S RW Spaces Tag displayed on the analyzer's display
Stream 1 statistics 1 62 Record 12 R
Simple Float 4 Maximum
Simple Float 4 Minimum
Simple Float 4 Average
Alarm relay 1 1 63 Record 24 RW
Simple Unsigned8 1 0
0 = Off
1 to 6 = Stream number
7 = Cleaning
Alarm source
Simple Unsigned8 1 1
1 = High process
2 = Low process
3 = High latch
4 = Low latch
5 = High annunciate
6 = Low annunciate
7 = Out of sample
Alarm type
Simple VisibleString 20 Alarm A ASCII string
Simple Unsigned8 1 1
0 = False
1 = True
Fail safe
Simple Unsigned8 1 1
0 = False
1 = True
Log enable
1
Store definitions – refer to page 7
2
Access definitions – refer to page 7
Description Slot
Slot
Index
Object
Typ e
Data Type Bytes Store
1
Access
2
Default Valid Range Note
Table B.1 Data Structure (Sheet 9 of 38)