Navigator 600 Series Analyzers
Silica and Phosphate Appendix B – Acyclic Parameter Mapping
IM/NAV6PBS Issue 1 17
Stream 2 Function Block Parameters
Stream 2 output 2 26 Record DS-33 5 D Cyclic / R
Output scale 2 28 Record DS-36 11 S RW
High range Simple Float 4 S RW 5000 ppb
SiO2 = 0 to 5000ppb
PO4 = 0 to 12ppm
P = 0 to 4ppm 
Low range Simple Float 4 S RW 0 ppb
SiO2 = 0 to 5000ppb
PO4 = 0 to 12ppm
P = 0 to 4ppm 
Units Simple Unsigned16 2 S RW ppb
1423 = ppm
1424 = ppb
1558 = mg/l
1559 = ug/l
1521 = ug/kg
1522 = mg/kg
Common units setting for all streams.
SiO2 = ppb, ug/l only
PO4 & P = ppm, mg/l & mg/kg only
Decimal point Simple Integer8 1 S RW 0 0 or 1
Alarm hysteresis 2 35 Simple Float 4 S RW 0
High-high alarm limit 2 37 Simple Float 4 S RW 5000 ppb
High alarm limit 2 39 Simple Float 4 S RW 50000 ppb
Low alarm limit 2 41 Simple Float 4 S RW 0 ppb
Low-low alarm limit 2 43 Simple Float 4 S RW 0 ppb
Simulate 2 50 Record DS-50 6 S RW
Affects only the Profibus output and 
not the value displayed on the 
analyzer's display 
Simulate status Unsigned8 1 S RW 0 x 80
Simulate value Float 4 S RW 0
Simulate enable/disable Unsigned8 1 S RW 0
0 = Disabled
Not 0 = Enabled
1
Store definitions – refer to page 7
2
Access definitions – refer to page 7
Description Slot
Slot
Index
Object
Typ e
Data Type Bytes Store
1
Access
2
Default Valid Range Note
Table B.1 Data Structure (Sheet 11 of 38)