EasyManua.ls Logo

ABB ScreenMaster RVG200 - Power Supply Connections

ABB ScreenMaster RVG200
16 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Accessing the recorder connection terminals
Referring to Fig. 3:
1. Press the terminal cover release plate in recess
A and
pull terminal cover
B away from the recorder body to
expose the terminal connections.
Power supply connections
Referring to Fig. 4:
1. Make connections to the power supply terminals
(module position E) as follows:
AC supply: terminals 1 (line), 2 (neutral), 3 (ground)
or
DC supply: terminals 4 (+), 5 (–)
Fig. 3 Accessing the recorder connection terminals
Note. Tighten terminal screws to a torque of
0.1 Nm (0.9 lbf.in).
Fig. 4 Power supply connections and fuse ratings
3
5
4
+
3
1
2
L
N
ABCD E
1
2
3
4
5
Warning. Use fuse rating 500 mA (max.) Type T
(UL category JDYX2)
Warning. Use fuse rating 2.5 A (max.) Type T
Neutral
Fuse 500 mA Type T IR = 100 A
Fuse
2.5 A Type T
AC Supply
DC Supply
100 to 240 V
50/60Hz
24VDC
Module positions
Power
supply
terminals
see below
Line
AC power supply
earth (ground)
stud
6 CI/RVG200–EN