TROUBLESHOOTING
The i-STAT Alinity is programmed to perform quality checks
throughout the testing cycle.
The instrument has several methods of notifying operators
of failed quality checks.
1. Quality Check Failures
- Are displayed when the instrument identies a problem
while running a cartridge or simulator
- There are 4 types of quality check failures:
1. Instrument
2. Cartridge
3. Sample
4. Software
- Screen displays the type of failure and instructions
for resolution
View Resoluon
Cause
Excess blood added to cartridge.
When filling this cartridge, the blood
advanced past the level indicated by
the 'fill to' arrow.
Help
?
Opons
Menu
Home
Cartridge Quality Check Failure
Cartridge Was Overfilled
Code: 37-01
Print
View
Entered Info
Opons
Menu
i-STAT CHEM8+
Pt: 654321
13OCT2025 08:33
View Cause
Cartridge Quality Check Failure
Resoluon
Cartridge Was Overfilled
When filling a cartridge, use care to
advance blood to the level indicated by
the 'fill to' arrow.
Repeat tesng with a freshly filled
cartridge.
Carefully observe the help provided
throughout the tesng pathway.
If the same quality check failure displays,
contact the system administrator for
further instrucon.
Code: 37-01
Help
?
Opons
Menu
Home
Print
View
Entered Info
Opons
Menu
i-STAT CHEM8+
Pt: 654321
13OCT2025 08:33