21
What It Means What To Do
2. The test
does not
start after
applying
the blood
drop.
Blood drop is too small 1. Add blood to the same sample area within
60 seconds, or repeat the test using a new
test strip and a larger blood drop.
Defective test strip
Sample applied after
meter turns o
1. Repeat the test using a new test strip. Wait for
and
to appear before applying blood or
control solution.
2. If problem persists, contact Customer Care.
Defective meter 1. Contact Customer Care.