22
FreeStyle Control Solution
Purpose
FreeStyle control solution is a red liquid that contains a xed amount of glucose.
The purpose of doing a control solution test is:
To practise testing without using your own blood
To conrm that the meter and test strips work together properly
When you are unsure of your blood glucose test results
Contact Customer Care for information on how to obtain control solution.
Control Solution Information
Use only FreeStyle control solution with the meter.
Replace the cap securely on the bottle immediately after use.
Control solution results are accurate only between 59° – 104° F (15° – 40° C).
Do not add water or other liquid to control solution.
There are three levels of control solution available (Low, Normal, High).
For information on how to obtain control solution, contact Customer Care.
IMPORTANT: Control solution results should fall within the control
solution range printed on the test strip vial. Do not use this range
when testing blood glucose.