EasyManua.ls Logo

Abbott FreeStyle Navigator - Page 84

Abbott FreeStyle Navigator
196 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Steps Do This
6. Set, review, or change your
Projected High glucose alarm.
a. Select Projected High from the Alarms menu.
•UsetheUP/DOWNArrowbuttonstohighlight
Projected High.
•PresstheRIGHT Option button to select
Projected High.
b. Review or change your settings.
•PresstheLEFT Option button (Next) to move
from one eld to the next.
•UsetheUP/DOWNArrowbuttonstochange
the setting in any eld.
•PresstheRIGHT Option button (Set) to accept the
settings and return to the Alarms menu.
Note: e receiver will respond with a brief sample
of the alarm you selected.
c. Decide what to do next.
Decide what settings you want:
• Alarmonoro.
• Sensitivity.
• Alarmtype(Setting).
– Low, medium, or high beep.
– Short, medium, or long
vibration.
•Youaresetting,reviewing,orchanging
all alarms.
Continue to the next step.
•Youarenishedsetting,reviewing,and
changing alarms.
Go to Step 11.
If en
If you want this much notice
before reaching your High
glucose level:
en set the
Sensitivity to:
•Maximumdetection(most
alarms, approximately 30
minutes in advance of events)
•Intermediatedetection
(intermediate alarms,
approximately 20 minutes in
advance of events)
•Lowdetection(fewestalarms,
approximately 10 minutes in
advance of events)
High
Medium
Low
Customer Care: 1-866-597-5520
74 Section 8
Set, Review, or Change the Alarm Settings
Alarm Settings 08:30A
Low Glucose (70)
High Glucose (275)
Projected Low (On)
Projected High (O)
Data Loss
Main Select
Alarm Settings 08:30A
Projected High Glucose
ON
Sensitivity: Low
Setting: Medium Beep
Next Set