IEC 60601 test level: 10 V/m; 80 MHz to 2.7 GHz; 80% AM at 1 KHz
Compliance level: 10 V/m; 80 MHz to 2.7 GHz; 80% AM at 1 KHz
Electromagnetic environment – guidance:
Immunity test: Proximity fields from RF wireless communications equipment; IEC
61000-4-3
IEC 60601 test level: See table below
Compliance level: Compliance to the tested levels
Electromagnetic environment – guidance: Portable RF communications equipment
(including peripherals such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the System, including cables
specified by Abbott Diabetes Care. Otherwise, degradation of the performance of
the System could result.
The table below lists the immunity test levels at specific test frequencies for testing
the e"ects of some wireless communications equipment. The frequencies and
services listed in the table are representative examples in healthcare and in various
locations where the System may be used.
ART42490-004_rev-B 4:05pm 03-May-2023

Other manuals for Abbott Libre 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott Libre 3 and is the answer not in the manual?

Abbott Libre 3 Specifications

General IconGeneral
Sensor TypeContinuous Glucose Monitoring (CGM)
ConnectivityBluetooth
Wear Duration14 days
CalibrationFactory calibrated
Mobile AppFreeStyle Libre 3 app
Warm-up Time1 hour
Readings FrequencyEvery minute
App CompatibilityiOS and Android
Operating Temperature10°C to 45°C
Storage Temperature4°C to 25°C
AccuracyMARD of 9.2%
Battery LifeSensor is disposable, no battery replacement needed

Related product manuals