* For glucose ≤69mg/dL (3.8mmol/L), the di"erences inmg/dL (mmol/L) are presented instead of relative
di"erences (%).
Table 4. Sensor accuracy over wear duration vs. YSI reference
Skin Interaction
Based on the examination of 101 study participants, the following incidence of skin
issues were observed. All skin issues were reported to be mild or moderate in
severity.
Bleeding – 3.0% of participants
Edema – 1.0% of participants