Energy Factors for BurstDR™ Stimulation Parameters
The following tables show energy factors according to various stimulation parameters for
BurstDR™ stimulation programs.
NOTE: Energy factors represent IPGs that provide 24 hours of daily stimulation using default
values for the Burst Frequency, Intra-burst Rate, and Pulse Width settings.
NOTE: In neurostimulation therapy, “dose” refers to the delivery of a quantity of energy to
tissue. A difference in “dose” in this context does not imply differences in expected
effectiveness response as it would with a drug. There is no demonstrated difference in safety
or effectiveness among these doses.
Table 10. Energy factors for various BurstDR stimulation parameters (350-ohm impedance)
Intermittent Dosage (On Time/Off Time)
0.2
47 31 25 23 16 13 7 77
0.4
71 43 34 28 20 17 10 89
0.6
93 53 41 34 24 20 12 104
0.8
117 65 49 40 29 25 13 118
1.0
146 76 56 46 34 28 16 134
1.2
175 106 84 73 50 42 24 240
1.4
202 121 95 82 56 48 26 270
1.6
230 135 106 91 63 55 29 301
1.8
247 151 120 100 70 60 33 330
2.0
279 163 133 110 76 65 35 361
Table 11. Energy factors for various BurstDR stimulation parameters (500-ohm impedance)
Intermittent Dosage (On Time/Off Time)
0.2
46 32 25 23 15 13 7 76
0.4
71 44 34 29 19 18 9 91
0.6
95 56 41 34 24 20 12 105
0.8
121 67 49 41 28 25 14 120
1.0
151 93 73 64 43 37 21 211
1.2
174 108 84 73 50 43 23 240
1.4
203 122 96 82 57 48 26 271
1.6
225 137 106 93 64 55 29 301
1.8
264 160 129 111 70 59 32 330
2.0
297 196 168 146 97 83 45 512