59
Your blood glucose may be extremely high or a meter or a test strip error has occurred.
If your test result matches how you feel, contact your healthcare professional immediately.
If your test result does not match how you feel, repeat the blood glucose test and see
Chapter 2, Unusual Blood Glucose Results.
If the E-3 code still appears for your blood glucose test, your blood glucose result may
be extremely high and above the system’s reading range. Contact your healthcare
professional immediately.
If the second test result does not match how you feel, perform a control test with the
control solution and a new test strip.
If the control result is within the acceptable range, review the proper testing
procedure and repeat the blood glucose test with a new test strip.
If the control result is not within the acceptable range, see Chapter 4,
Understanding Out-of-Range Control Results.
The code chip is incorrect. Turn the meter off and insert a new code chip.
50705_05853397002_EN.indb 59 4/24/12 8:32 AM