011807-EN--B (2022-12_S617 S717F S900 S950 C50_IFU)
11 / 49
3.5.6 DOCKING STATIONS
The MINI DOCK U_USB2 module is built-in docking station for SOPRO 617, SOPRO 717FIRST, SOPROCARE and SOPROLIFE
cameras.
The MINI DOCK U_USB2 docking station is to be powered directly by a USB 2.0 compatible port.
The docking station is turned ON/OFF according to the computer electrical supply. In case of use of a handpiece holder, connect
the strap on the ON/OFF selector (cf. installation diagram).
Its compact size and light weight allow an easy installation into different dental units. This product is composed of a docking
station and a linking cable to the handpiece.
The electrical supply for all cameras is directly performed through the computer USB port. The voltage powering the camera
is of continuous 5 V low voltage type (0.5 A).
3.5.7 SINGLE-USE PROTECTIVE SHEATH
A single-use protective sheath is put on the intraoral camera’s head as a hygienic protection between each patient.
3.5.8 INTRAORAL TIPS (C50TIPS AND SOPROTIPS)
C50TIPS and SOPROTIPS intraoral tips are respectively used with C50 and SOPROCARE / SOPROLIFE intraoral cameras when
using their DIAGNOSTIC AID mode.
Intraoral tips are placed on the camera head (itself protected by the protective sheath) to improve fluorescence imaging quality
by shielding the camera from ambient light.
3.6 USER POPULATION RECOMMENDATIONS
3.6.1 USER POPULATION
The devices are intended for use by:
• A qualified and certified dentist
• A qualified and certified dental hygienist
• A qualified and certified nurse
• The following requirements must be fulfilled by the user: Any vision problems must be corrected by glasses or contact
lenses / Colour perception is mandatory
• The user must wear gloves
The device is not intended for self-treatment.
The device can be used by any adult practitioner of any weight, age, height, gender and nationality.
3.6.2 SPECIFIC USER TRAINING
No specific training other than initial professional training is required to use this medical device.
3.6.3 DEVICES IN CONTACT WITH THE USER
The dental cameras and accessories associated are in indirect contact with the user, who wears gloves during use of the device.
The dental cameras do not enter in direct contact with the patient’s mouth.
The protective sheaths, C50TIPS and SOPROTIPS are in direct contact with the patient mouth and are biocompatible according
to the ISO 10993-1:2018 standard.
3.7 PATIENT POPULATION RECOMMENDATIONS
3.7.1 PATIENT POPULATION
This medical device is designed to be used with the following patient populations:
• Children
• Adolescents
• Adults
• Elderly persons
This medical device can be used irrespectively of the patient's details such as weight, age, height, gender, and nationality.