Guide and declaration of the manufacturer - electromagnetic immunity
The standard device « Sopix² dental digital x-ray system » is intended to be used in the electromagnetic
environment specified below. The user should make sure it is used in this environment
Immunity trial
Electrostatic discharges
EN 61000-4-2
Far transient bursts
EN 61000-4-4
Voltage shocks
EN 61000-4-5
Dips, brief outages and
power voltage variation
EN 61000-4-11
Magnetic field with the
network frequency
(50/60 Hz)
± 6 kV when
in Contact
± 8 kV in the
air
± 2 kV for the
feed cables
± 1 kV for the
input/output
cables
Differential
mode ± 1 kV
Common
mode ± 2 kV
• <5% U
T
- for
10 ms
• 40% U
T
- for
100 ms
• 70% U
T
- for
500 ms
• <5% U
T
- for
5 s
3 A/m
± 6 kV
± 8 kV
± 2 kV
± 1 kV
± 1 kV
N.A.
<5% U
T
10 ms
<40% U
T
100 ms
<70% U
T
500 ms
<5% U
T
5 s
3 A/m
The floor should be wooden, concrete or tile. If
the floor is covered with a synthetic material, the
relative humidity should be at least 30%.
The main power supply quality should be the one
of a traditional commercial or hospital
environment.
The main power supply quality should be one of a
traditional commercial or hospital environment.
The main power supply quality should be one of a
traditional commercial or hospital environment.
If the user of the standard device « Sopix² dental
digital x-ray system » requires it to continue to
operate during main power supply outages, it is
recommended the standard device « Sopix² dental
digital x-ray system » is fed by an inverter or a
battery.
The magnetic field with the network frequency
should be at a characteristic level of a location in
a traditional commercial or hospital environment.
CEI 60601
Severity level
Compliance
level
Electromagnetic environment
Guide
Note: UT is the power voltage nominal value applied during the trial.