Sopix
²
• User manual
ENGLISH
19
MAINTENANCE
The Sopix² digital x-ray system does not need any maintenance if it is used according to the manufacturer use
and cleaning instructions.
Before first using any device from the Sopix² range, it is imperative to follow the complete disinfecting
procedure.
Any device from the Sopix² range returned to the manufacturer or distributor for servicing or maintenance
should have been carefully disinfected before shipping.
Any device of the Sopix² range coming back from servicing or maintenance operation should follow the complete
disinfecting procedure before being used.
8.1. CONTROLLER MAINTENANCE
DESCRIPTION
Decontamination
and disinfecting.
WARNING
Do not use products
containing:
• Ammoniac,
trichloroethylene
• Dichloroethylene
• Ammonium
hydrochlorid
• Chlorinated and
aromatic
hydrocarbon
• Ethylene
dichloride
• Methylene
chloride
• Ketones
Deterioration risks
of the plastic parts.
Do not
scrub.
Do not
rinse.
RECOMMENDATIONS
• Surface disinfectant
spray of Septol™
type Spray Surface
without Aldehyde –
Pierre Rolland.
• Surface cleaning
and disinfecting
wipes of Wipes
Septol™ type -
Pierre Rolland.
Spray at 40 cm
from the surface
and allow to dry.
Take the wipe,
wring it, and then
scrub until
obtaining visible
cleanliness.
Allow to dry in the
open air.
Carefully close the
packaging box.
USE INSTRUCTIONS AND PRECAUTIONS