EasyManua.ls Logo

Acteon SOPROCARE - Page 13

Acteon SOPROCARE
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ENGLISH
SOPROCARE • User’s manual
12
Guide and declaration of the manufacturer - electromagnetic immunity
SOPRO device is intended to be used in the electromagnetic environment speci ed below.
The user should make sure it is used in this environment.
Immunity trial
IEC 60601
Severity level
Compliance
level
Electromagnetic environment
Guide
Electrostatic
discharges
EN 61000-4-2
± 6 kV when in contact
± 8 kV in the air
± 6 kV
± 8 kV
The oor should be wooden, concrete or tile. If the
oor is covered with a synthetic material, the relative
humidity should be at least 30%.
Far transient bursts
EN 61000-4-4
± 2 kV for the feed cables
± 1 kV for the input/out-
put cables
± 2 kV
± 1 kV
The main power supply quality should be one of a
traditional commercial or hospital environment.
Voltage shocks
EN 61000-4-5
Differential mode ± 1 kV
Common mode ± 2 kV
± 1 kV
N.A.
The main power supply quality should be one of a
traditional commercial or hospital environment.
Dips, brief outages
and power voltage
variation
EN 61000-4-11
• <5% Ut - for 10 ms
• 40% Ut - for 100 ms
• 70% Ut - for 500 ms
• <5% Ut - for 5 s
<5% Ut 10 ms
<40% Ut 100 ms
<70% Ut 500 ms
<5% Ut 5 s
The main power supply quality should be one of a
traditional commercial or hospital environment.
If the user of SOPROCARE device requires it to continue
to operate during main power supply outages, it is
recommended SOPROCARE device is fed by an inverter
or a battery.
Magnetic eld with the
network frequency
(50/60 Hz)
3 A/m 3 A/m
The magnetic eld with the network frequency should
be at a characteristic level of a location in a traditional
commercial or hospital environment.
Note: Ut is the power voltage nominal value applied during the trial.

Related product manuals