DESCRIPTION OF THE CONNECTION BOXES
WARNING:
Devices that connect to the inputs / outputs must conform to the IEC 60950-1 standard. Risk of electrical shock.
9.1. DOCK M_USB2 AND DOCK M_VIDEO POWER SUPPLY
SOPROCARE power supply is connected to the power outlet. The other end of the cord is connected to the connection box where the
symbol 6 V is located (PHIHONG, PSA 10R-060 Model or FRIWO, MPP15 FW 7555M/06 model).
The power supply automatically adapts to the electric networks 115 V~ - 230 V~; 60 Hz - 50 Hz; 0.5 A. SOPROCARE voltage is powered
by 6V
of a continuous low voltage type.
9.2. DOCK MU_USB2 / DOCK MU_VIDEO POWER SUPPLY
The electrical connection of this connection box should be performed by the installer. SOPROCARE power supply is powered through
the connection box that should be connected to 24 V~; 50 Hz - 60 Hz; 10 VA.
9.3. DOCK U_USB2 POWER SUPPLY
The electrical connection of this connection box should be performed by the installer. SOPROCARE power supply is powered through
the connection box that should be connected to 24 V~; 50 Hz - 60 Hz; 15 VA.
9.4. DOCK USB2 POWER SUPPLY
The dental camera electrical supply is directly powered through the computer USB port. The voltage powering the camera is of
continuous 5 V low voltage type (0.5 A).