3LP700 Rev10
(4C35/4D35/3255/MK05 Service Manual)
Page 5 of 5
setscrews (Figure 1) in the clapper,
then reposition the stir wire so that it
is resting on the same level as the end
of the sample probe (Figure 5).
c. Firmly retighten the stir/freeze wire
setscrew(s).
Make sure the stir wire
does not rub on the
sides of the slot in the
mandrel.
Stir wire should be parallel to
sample probe, and the space
should be about .06 inches
(1.5 mm).
Bottom of sample
probe and end of
stir wire should rest
on this level.
This level is not used with
cryoscopes.
Mandrel
Raised V
Centerline
Alignment
Tool
Sample Probe
Stir Wire
Figure 5. Cryoscope Alignment