configuration, you will need to enter your
configuration settings and recalibrate your
instrument (next step).
If you are using the original processor
board (application board replacement),
you will not need to reenter any settings
and may skip to step 8.
6. Use the SETUP menu to reset the sample
and block probe bin numbers and the
sample stir amplitude, and any other cus-
tomized settings.
Note: If values are unavailable, refer to the
Users Guide for setup procedures.
7. Reset the serial number (see the following
procedure).
8. Press STOP to exit the Setup menu. The
instrument will run the self-diagnostics
program.
9. Recalibrate the instrument. For further
information, refer to your Users Guide.
10. If you have upgraded from a single con-
trol board to the two-board set, see the
Supplemental Information document.
11. If you have upgraded from a single con-
trol board to the two-board set and your
instrument has the black plastic keyswitch
located on the rear of the instrument, your
Operator and Supervisor positions will
have changed. Replacement labels have
been provided along with this upgrade.
Place the “OPERATOR” label along the
top of the keyswitch, and place the
“SUPERVISOR” label along the right
side.
Reset Serial Number
1. If there is a Supervisor/Operator
keyswitch, turn it to the Supervisor posi-
tion (if required).
2. At “Press START to Continue”, press
SETUP.
3. Cycle through the menu options by press-
ing < or > until you reach “Serial
Number”, and then press START. The
display will read “Serial#:********”.
4. Press the following keys consecutively:
2, 4, CLEAR. The displayed serial num-
ber will disappear.
5. Enter each digit of the serial number as it
appears on the label on the back of the
instrument (Note: the letter suffix at the
end of the serial number will not be
entered). When the number is correct,
press ENTER.
The serial number will be recorded in
memory and should appear the next time
you power up your instrument.
4D3P620 Rev3
(4D35 Service Manual)
Page 4 of 4
SW1
1 2 3
3D3 ON OFF OFF
MK3 ON ON OFF
4D3 OFF OFF OFF
MK2 OFF ON OFF
Program N/A N/A ON
NOTE:
ON = 1 = CLOSED
OFF = 0 = OPEN
XXX620 Option Switch Setup