75
Product Specifications
Appendix B
Electrical
Power Requirement: 100 to 240 VAC (50/60 Hz)
Fuses (2): 250V time delay (Type T): 2-Amp
Power Consumption: 95 Watts
Memory Backup: Integral lithium cell; 10-year life (typical); (not user-
replaceable)
Sample Volume: 2.0 to 2.5 mL
Sample Capacity: Single sample
Readout: 20-character vacuum fluorescent display
Units: mºC or mºH
Range: 0 to 1000 mºC or mºH
Resolution: 1 mºC or mºH
Communications: On-board printer, DTE EIA-232/V.24 (RS-232) serial port
and optional barcode scanner
Performance at
Reference Conditions
1
Linearity: Less than 0.5% from a straight line
Repeatability: Std. Deviation < 2 mºC or mºH between 0 and -400 mºC
Std. Deviation < 0.5% of value between -400 and -1000 mºC
Drift: Less than 1 unit per month
Performance Over
Operating Conditions
Temperature Less than 1 unit for every 5°C (9°F) ambient
Effects: temperature change.
1
Reference Conditions: 20°C to 25°C (68°F to 77°F); 40% to 60% Relative Humidity; tolerances of reference
or calibration solutions excluded.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments Advanced Cryoscope 4250 and is the answer not in the manual?

Advanced Instruments Advanced Cryoscope 4250 Specifications

General IconGeneral
BrandAdvanced Instruments
ModelAdvanced Cryoscope 4250
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals