EasyManua.ls Logo

Advanced Instruments Fluorophos FLM300 - Chapter 1 FLM300 Overview; Removing the Cover; Cleaning

Advanced Instruments Fluorophos FLM300
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7
Fluorophos® ALP Test System Model FLM300 Service Manual
Chapter 1
FLM300 Overview
Shut o and unplug the FLM300 before
performing the following procedures.
1.1 Removing the Cover
Tools needed:
Phillips head screwdriver
1. Remove the four screws from the sides of the
FLM300.
2. Remove the heating block from the top of the
instrument.
NOTE: Lift the cover o and place it on its side to the left
of the instrument. There are several cable
connections between the application PCB
mounted in the base of the instruments and the
cover. If required, these cables can be
disconnected after the cover is next to the
instrument.
Figure 1: Side View of FLM300
1.2 Cleaning
The FLM300 should be cleaned on a regular basis.
Performance problems are often the result of dirt and
residue that build up in the cuvette chamber.
Cuvette Holder Cleaners are used to clean residue and
small spills from the cuvette chamber and the separate
cuvette heating block. The directions are as follows:
1. Moisten the cleaner, insert it into the chamber and
leave it for several minutes.
2. Gently pull the cleaner out and discard. (Cuvette
Holder Cleaners are not reusable).
If the Cuvette Holder Cleaners cannot suciently absorb
a spill, the FLM300 must be disassembled and
thoroughly cleaned following the procedure below:
Tools needed:
Pliers
Flat-bladed screwdriver
Phillips screwdriver
1. Disconnect the Photomultiplier Tube, cuvette, and
sample illumination LED cables from the
application PCB ( , [A] [B] [C]).
2. Remove the six (6) screws that secure the optical
bench to the base.
3. Remove the excitation filter holder and the
emission filter holder.
4. Remove the two (2) screws from the optical bench
and cuvette holder ( , [D]).
5. Pull the cuvette holder out from the optical bench
( , [E]).
6. Remove the two (2) screws from the mirror access
plate ( , [F]). Remove foam pad that secures the
mirror ( , [G]).
7. Wipe all parts of the optical assembly, including
the Photomultiplier Tube, with non-abrasive
tissues. For dicult spills, use small amounts of
water and soap. (To disassemble the
Photomultiplier Tube, see Photomultiplier Tube
(PMT) and PMT Socket Replacement in the
Replacement Procedures.)
Do not use water on wiring or other electronic
components. Do not immerse any parts in water.

Related product manuals