3
Fluorophos® ALP Test System Model FLM300 Service Manual
Safe use
The FLM300 uses AC voltage. Before removing the cover, make sure the instrument is turned off and unplugged.
To reduce the risk of bodily injury, electric shock, fi re, and damage to your instrument, please read and
observe the following precautions.
If the product is used in a manner not in accordance with the equipment design, operating instructions or manufacturer’s
recommendations, the operation of the product may be impaired to the extent that a safety hazard is created.
Do not attempt to perform electrical work if you are not fully qualifi ed. This manual is not a substitute for electrical
training.
General Cautions
This product should be operated only with the type of
power source indicated on the product’s electrical
ratings label. Refer to the installation instructions
included with the product.
If the power cord provided is replaced for any reason or
if an alternate cord is used, the cord must be approved
for use in the local country. The power cord must be
approved for the product’s listed operating voltage and
be rated at least 20% greater than the ampere ratings
marked on the product’s electrical ratings label. The
cord end that connects to the product must have an IEC
60320 connector.
Plug the product into an approved grounded electrical
outlet.
Do not disable the power cord’s grounding plug.
If an extension cord or power strip is used, make sure
that the cord or strip is rated for the product, and that
the total ampere ratings of all products plugged into
the extension cord or strip do not exceed 80% of the
cord’s or strip’s rating limit.
Route power cords so that they will not be walked on,
tripped on, or pinched by items placed upon or
against them. Pay particular attention to the plug,
electrical outlet, and the point where the cord exits the
product.
Do not pull on cords and cables. When unplugging
cords or cables, grasp the corresponding connector.
Do not install or use this product in any area subject to
extreme short-term temperature variations, or
locations that exceed the specifi ed operating
environment temperatures.
Never use this product in a wet area.
To avoid injury or fi re hazard, do not operate this
product in an explosive atmosphere.