EasyManua.ls Logo

AEMC 1000N - Specifications

Default Icon
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Megohmmeter Model 1000N/1100N
- 7 -
6SHFLILFDWLRQV
INSULATION TESTS
DC Test Voltages:
50, 100, 250, 500, 1000V
Megohm Ranges:
For each test voltage two direct reading ranges:
1 to 1000M
1 to 1000G
Short Circuit Current:
5mA (max)
Accuracy:
5% of reading typical
15% of reading between 100 and 1000G
on the 50V range
Charging Time:
1000 M
range: 0.3 seconds/µF
1000 G
range: 4 seconds/µF
Discharging Time:
Automatic discharge when test button is released; 0.1 seconds/µF
Scale:
Two large overlapping scales: 4.7" (119 mm) for each range
Test Voltage Generation:
Solid state circuitry generating rated test voltage across the full range
RESISTANCE TESTS
Test Current:
Constant 1mA DC
Kilohm Range:
1 to 1000k
(1M
)
Maximum Test Voltage:
1200V
DC
Accuracy:
5% of reading typical
CONTINUITY TESTS
Resistance Range:
0.1 to 100
; 3
mid-scale
Accuracy:
6% of reading
Audible Tone:
Continuous for continuity resistances under 200
Fuse:
High interrupting capacity (0.315A)