Digital Megohmmeter Model 1026
6
Electrical Specifications:
AC Voltage
Range Resolution Accuracy
600V 0.1V ± (1.5% rdg + 5 dgts)
DC Voltage
Range Resolution Accuracy
600V 0.1V ± (1% rdg + 3 dgts)
Ohms (Autoranging)
Range Resolution Accuracy
400 0.1
± (1% rdg + 5 dgts)
4000 1
Continuity beeper
Range Active Protection
40
600Vrms
M (Autoranging)
Range Resolution Accuracy
4/40/400/4000M (250V)
1k
± (3%rdg + 5dgts
< 2000M)
± (5%rdg + 5dgts
2000M to 4000M)
4/40/400/4000M (500V)
4/40/400/4000M (1000V)
Voltage Output: -10% to +20% of selected range
Short Circuit Current: 250/500/100V: 2mA max
Safety Specifications:
IEC 1010, Cat. III, 600V, Pollution Degree 2
Altitude up to 2000 meters (6000 ft.)
Indoor use only
EN50081-1; EN50082-1
www.GlobalTestSupply.com