EasyManuals Logo

AEMC 1026 User Manual

AEMC 1026
16 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #8 background imageLoading...
Page #8 background image
Digital Megohmmeter Model 1026
6
Electrical Specifications:
AC Voltage
Range Resolution Accuracy
600V 0.1V ± (1.5% rdg + 5 dgts)
DC Voltage
Range Resolution Accuracy
600V 0.1V ± (1% rdg + 3 dgts)
Ohms (Autoranging)
Range Resolution Accuracy
400 0.1
± (1% rdg + 5 dgts)
4000 1
Continuity beeper
Range Active Protection
40
600Vrms
M (Autoranging)
Range Resolution Accuracy
4/40/400/4000M (250V)
1k
± (3%rdg + 5dgts
< 2000M)
± (5%rdg + 5dgts
2000M to 4000M)
4/40/400/4000M (500V)
4/40/400/4000M (1000V)
Voltage Output: -10% to +20% of selected range
Short Circuit Current: 250/500/100V: 2mA max
Safety Specifications:
IEC 1010, Cat. III, 600V, Pollution Degree 2
Altitude up to 2000 meters (6000 ft.)
Indoor use only
EN50081-1; EN50082-1
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com