microAeth
®
MA Series MA200, MA300, MA350 Quick Start Guide
AethLabs
26
Code Value
Readable
status
Status Description
32768 Tape not ready
Tape is not ready. Check that the tape is installed
and clamped.
65536
Tape transport
not ready
Tape transport is not ready. Check that tape is
installed and clamped and there are no other errors.
131072 Ext. power
Device is powered through the barrel jack input or
circular connector (MA350) by external 5v source.
262144 Invalid date/time
Date and time are not set or are invalid. Set the
clock with proper date/time.
524288 Tape error
Tape Error. A generic error for the tape. Possible
reasons are tape is not detected in the instrument.
16777216
WiFi Forced
Timebase to 60s
When WiFi is enabled and timebaseis set to less
than 60 seconds, the timebase will be forced to 60
seconds as this is the minimum timebase for WiFi
functionality.
4294967296 WiFi Line Full WiFi data buer is full.
274877906944
Remote power
down
Power down command received remotely via serial
connection.