Registering the DR Detector on another NX
Workstation
The DR Detector can be used for examinations on different NX workstations.
The DR Detector is set up to communicate to a specific NX workstation. The
procedure of registering the DR Detector to another NX workstation switches
the availability of the DR Detector between NX workstations.
In a configuration with multiple NX workstations sharing a DR Detector, each
NX workstation is equipped with a DR Detector connector cable (wired
configuration) or a DR Detector registration cable (wireless configuration).
To register the DR Detector and make a connection to another X-ray room:
1. Make sure that no empty thumbnail is selected in the NX workstation, so
no other DR Detector is active.
2. Connect the DR Detector to the NX workstation using the DR Detector
Cable.
Wait for the DR Detector Switch to display the wired connection icon.
The DR Detector is set up to make connection to the selected NX
workstation.
3. To operate wireless, disconnect the DR Detector registration cable.
In specific configurations, the DR Detector turns to error status directly after
registration to another NX workstation, or after attempting a first exposure. In
this case, the DR Detector must be started again by removing the battery and
inserting it again.
Related Links
Starting the DR Detector (wireless configuration) on page 74
Stopping the DR Detector (wireless configuration) on page 92
DR 10e, DR 14e, DR 17e | Advanced Operating | 107
0370C EN 20200228 1449