Electromagnetic immunity
This DR Detector is intended for operation in the electromagnetic
environment given below. The user of the DR Detector should ensure that it is
used in such an environment.
Immunity Test
IEC 60601-1-2
Test Level
Compliance
Level
Electromagnetic en-
vironment - guid-
ance
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV con-
tact
± 15 kV air
Floors should be
wood, concrete or ce-
ramic tile. If floors
are covered with syn-
thetic material, the
relative humidity
should be at least
30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 2 kV for
power supply
lines
± 1 kV for in-
put/output
lines
Mains power quality
should be that of a
typical commercial or
hospital environ-
ment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV
line(s) to
line(s)
± 2 kV
line(s) to
earth
Mains power quality
should be that of a
typical commercial or
hospital environ-
ment.
Voltage dips,
short interrup-
tions and voltage
variations on
power supply in-
put lines.
IEC 61000-4-11
0% U
T
for 0.5 cy-
cle
0% U
T
for 1 cycle
70% U
T
(30% dip
in U
T
) for 25 cy-
cles
0% U
T
for 5 s
0% U
T
for 0.5
cycle
0% U
T
for 1
cycle
70% U
T
(30% dip in
U
T
) for 25 cy-
cles
0% U
T
for 5 s
Mains power quality
should be that of a
typical commercial or
hospital environ-
ment. If the user of
the device requires
continued operation
during power mains
interruptions, it is
recommended that
the device be pow-
ered from an uninter-
ruptible power supply
or a battery.
DR 10e, DR 14e, DR 17e | Remarks for HF-emission and immunity | 127
0370C EN 20200228 1449