Use of protective plastic envelope
WARNING:
Liquids ingressing the DR Detector may cause malfunction and
contamination.
If there is a chance that the detector comes in contact with liquids (bodily
fluids, water, parenteral nutrition,...) or in direct contact with patient or
patient tissue, the DR Detector must be wrapped in a protective plastic
envelope while performing the examination.
It is considered good clinical practice to use a single-use protective envelope in
all cases where contact of the device or contaminants is expected, to avoid
contamination of others.
Make sure that the plastic envelope is not wrinkled to avoid the creases
showing in the image.
36 | DR 18M, DR 24M | Introduction to the DR Detector
0380A EN 20180703 1429