Image shows artifacts
Details An exposure is performed and the image on the NX
workstation shows lines and banding artifacts.
Cause The exposure time exceeded the integration time of the
DR detector (configured during installation).
Brief Solution
Depending on the frequency of the problem, take
following measures:
1. If this situation is exceptional and can be justified by
the high compression thickness or by the type of
examination, try to reduce the exposure time on the
X-ray system.
2. Perform a typical flat field exposure to check the
general image quality.
3. Contact your local service representative, who can
extend the integration time of the DR detector.
Related Links
Daily inspection on page 41
DR 18M, DR 24M | Problem solving | 65