< 5% Ur (95%
breakthrough of
Ur) for 5 s
< 5% Ur (95%
breakthrough of
Ur) for 5 s
free of interruptions
or a battery.
Magnetic field at
the supply
frequency
(50/60 Hz) in
accordance with
IEC 61000-4-8
3 A/m 3 A/m Magnetic field at the
network frequency
should correspond to
the typical values as
they are in a
commercial and
clinical environment.
Note: Ur is the alternating voltage.
Tests of Resistance
to Disruption
IEC 60601
Test Level
Level of
Agreement
Electromagnetic
Environment
Use portable and mobile
radio sets at a safe
distance from the DR
Detector (including the
leads) not closer than the
recommended protective
distance, which is
calculated according to
the equation suitable for
the transmission
frequency.
Recommended protective
distance:
Conducted high
frequency
disturbance
variables in
accordance with
IEC 61000-4-6
3 V
eff
150 kHz to
80 MHz
3 V
eff
150 kHz to 80
MHz
d = 1.2
Radiated high
frequency
disturbance
variables in
accordance with
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
d = 1.2 80 MHz to
800 MHz
76 | DR 18M, DR 24M | Remarks for HF-emission and immunity
0380A EN 20180703 1429