Information message Description
Max mAs Maximum mAs = 500 mAs (regulatory limit for
AEC)
Min ms Minimum exposure time (generator limit)
Max ms Maximum exposure time (generator limit)
Min ms & Min mA Minimum exposure time and minimum mA
(generator limit)
Max ms & Max mA Maximum exposure time and maximum mA
(generator limit)
Max Power Maximum power (generator limit)
Space Charge Filament emission limit for a combination of kVp
and mA in the selected focal spot.
Max Energy (60kJ) The maximum energy cannot exceed 60 kJ (kVp x
mAs = 60 kJ) (regulatory limit for AEC).
Instant Power Instantaneous power limit of the X-ray tube
depending on the anode temperature, the selected
exposure time and the selected focal spot.
50 | DX-D Software Console | Problem solving
0189D EN 20180329 1421