Indicator Description
TECHNIQUE ERROR Caution: If “TECHNIQUE ERROR” is
displayed during exposure it means
the exposure has been interrupted by
the “Security Timer” because of a
system failure. You should call field
service.
This error can also be shown if a failure on the
automatic collimator has been detected (blades are
full open or in movement during exposure, etc.). In
this case the indicator lights up continuously.
PANEL OFF Indicates that the detector panel has been
automatically powered off due to:
a power off in the X-ray generator control.
an overheat (above 47 °C) in the detector panel.
Exposure with the DR detector is inhibited
except for the free exposure mode.
(G)E63 Excessive current in auxiliary winding during
acceleration to 10000 RPM. Retry the exposure.
52 | DX-D Software Console | Problem solving