Chapter 3: Testing Logic Analyzer Performance
Testing the module using a 16700-series Mainframe
c Select the Acquisition Depth field, then select 128K.
d Select the Trigger Position field, then select Start.
e Select the Count field, then select Off.
3 Configure the Format settings.
a In the logic analyzer Setup and Trigger window, select the Format tab.
b Under the Format tab, select Pod Assignment.
c In the Pod Assignment window, use the mouse to drag the pods from
the Analyzer 2 column to the Analyzer 1 column.
d Select Close to close the Pod Assignment window.
e Under the Format tab, use the scroll bar to scroll to the fields for Pod
being tested (Pod 1).
f Select the field showing the channel assignment for the pod being
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com