Issued: 01-05-2012 Revision: 1.0 Copyright 2012 Agilent Technologies
Page 7 of 13
meter self-test and pH, ORP, ISE, DO and conductivity functionality.
For detailed calibration procedures refer to the User Guide of 3200 series meters.
3.1 Meter Power Supply
Check the power adapter input and output.
They should meet: Input 240 VAC, 1 A
Output 9VDC, 800mA
3.2 Meter Self-Test
3.2.1 Temperature test for all meters
Connect the ATC temperature diagnostic tool (PN:5185-8390) to the Temp. socket of
meters;
Turn on the meter to enter into measurement state.
The meter should display a temperature reading between 49.0 to 51.0 °C.
3.2.2 Check the potential (mV) value for 3200P/3200I/3200M
Disconnect the pH electrode.
Connect the short circuit plug (PN:G4383-40000) to the pH socket.
Turn on the meter.
Start measuring pH.
The meter should display a potential (mV) reading between -0.5 to 0.5 mV (also
called mV Zero).
3.2.3 Check pH value for 3200P/3200M
Disconnect the ATC probe if plugged.
Set the temperature to 25 °C manually.
Connect the short circuit plug (PN:G4383-40000) to the pH socket.
When the potential (mV) is zero, the meter should display pH 7.00 (±0.01).
3.2.4 Check pX value for 3200I/3200M
Disconnect the ATC probe if plugged.
Set the temperature to 25 °C manually.
Connect the short circuit plug (PN:G4383-40000) to the pH socket.
When the potential (mV) is zero, the meter should display pX 2.00 (±0.01).
3.2.5 Check for current Response for 3200D/3200M
Connect the meter with the ATC temperature diagnostic tool (PN:5185-8390) shipped
with the meter.
Turn on the meter. Enter the measurement state.
The meter should display a current between 680 to 720 nA.
When the diagnostic tool is unplugged, the meter should display a current from 0 to 5
nA.
3.2.6 Check conductivity for 3200C/3200M
Disconnect the ATC temperature diagnostic tool if plugged and manually set the
temperature to 25°C.
Connect the meter with conductivity diagnostic tool (PN:5185-8391) shipped with the
meter.
Turn on the meter. Enter the measurement state.
It will display:
Conductivity: nearly 100 μS/cm

Other manuals for Agilent Technologies 3200 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 3200 Series and is the answer not in the manual?

Agilent Technologies 3200 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model3200 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals