4339B
To
Use Comparator
Function
Setting
the
Limit
V
alues
1.
Press
2. Enter
the lower
limit value
using
the
numeric
ENTRY
keys
,
then
press
to enter
the
value
.
Y
ou
can
set
the
value
from
0
9.900
2
10
37
to
9.900
2
10
37
.
3.
Press
.
4.
Enter
the
upper
limit
value
using
the
numeric
ENTRY
keys
,
then press
to
enter
the
value
and
to
exit.
Y
ou
can
set
the
value
from
0
9.900
2
10
37
to
9.900
2
10
37
.
Sorting
T
o
start
sorting,
Press
.
The
Comprtr
On
annunciator(
9
)
turns
ON.
T
o
turn
sorting
OFF
,
Press
again.
The
Comprtr
On
annunciator
turns OFF
.
The
sorting results
are HIGH,
IN,
and
LOW
.
Where
,
HIGH
greater than
higher limit
IN
between higher
limit and
lower limit
LOW
less than
lower
limit
The 4339B shows the comparison results using the display
, beeper
, printer
,and
16064B LED Display/Trigger Box
.(T
o use the 16064B
, see \A
ccessories A
vailable" later in this
chapter.)
For result output to the display
,see\T
o Select Display Mode
" in the next page
.
For result output to the beeper
, see \T
o
Select Beeper Mode
" in the next page
.
For result output to the printer
, see \T
o
Print Measurement Data
" later in this chapter
.
Operating the 4339B 2-11

Table of Contents

Other manuals for Agilent Technologies 4339B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4339B and is the answer not in the manual?

Agilent Technologies 4339B Specifications

General IconGeneral
BrandAgilent Technologies
Model4339B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals