EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>4339B

Agilent Technologies 4339B User Guide

Agilent Technologies 4339B
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #40 background imageLoading...
Page #40 background image
4339B
h. Press .
i.
Select
OpenMeas
using
or and
press
.
After
a
while
, the
OPEN correction
will be
completed
with
the
message
\
Correction
Complete
".
(If
Out
Of
Limit
is
displayed,
see
\
T
o
P
erform
OPEN
Correction
|Canceling
the
stray
admittance
in
parallel
with
the
DUT
"
in
Chapter 2.)
j.
Press
to
turn
OFF
the
test
voltage
.
The
V
Output
indicator
will turn
OFF.
6.
Connect
the DUT
and close
the cover
.
7.
Press
.
8.
Select
R
using
or and
press
to
select
the
resistance
measurement
mode
.
R
(Resistance).
9.
Set
the
measurement
sequence
charging time
to 1
minute.
a.
Press
.
b. Select
Chrg
using
or and press .
c. Press .
d. Select
Exit
and press to exit.
3-4 Measurement Examples

Table of Contents

Other manuals for Agilent Technologies 4339B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4339B and is the answer not in the manual?

Agilent Technologies 4339B Specifications

General IconGeneral
BrandAgilent Technologies
Model4339B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals