4339B
h. Press .
i.
Select
OpenMeas
using
or and
press
.
After
a
while
, the
OPEN correction
will be
completed
with
the
message
\
Correction
Complete
".
(If
Out
Of
Limit
is
displayed,
see
\
T
o
P
erform
OPEN
Correction
|Canceling
the
stray
admittance
in
parallel
with
the
DUT
"
in
Chapter 2.)
j.
Press
to
turn
OFF
the
test
voltage
.
The
V
Output
indicator
will turn
OFF.
6.
Connect
the DUT
and close
the cover
.
7.
Press
.
8.
Select
R
using
or and
press
to
select
the
resistance
measurement
mode
.
R
(Resistance).
9.
Set
the
measurement
sequence
charging time
to 1
minute.
a.
Press
.
b. Select
Chrg
using
or and press .
c. Press .
d. Select
Exit
and press to exit.
3-4 Measurement Examples