Manual Change
Source Voltage Accuracy Test
Step 28. Press [Trig] to measure.
Step 29. Record the 4339B reading on the calculation sheet.
Step 30. Press [V Output] to turn the source voltage OFF. (V Output indicator turns OFF.)
Step 31. Calculate the test result according to the calculation sheet, and record the result into the
performance test record.
Step 32. Perform this test for all settings listed in Table 5. The source voltage must be turned OFF
after each test.
Table 5 Resistance Measurement Accuracy Test Settings (Grounded)
Standard
Resistor
Voltage
Setting
16353D 100V
16353H 10V