EasyManua.ls Logo

Agilent Technologies 4339B - Page 6

Agilent Technologies 4339B
18 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Manual Change
Source Voltage Accuracy Test
Figure 1 Source Voltage Accuracy Test Setup
Step 3. Press the 3458A Multimeters [DCV] to set the measurement mode to DC voltage.
Step 4. Set the V Measurement Adapters ID switches and Output switch as follows:
Step 5. On the 4339B, press [V Output] to turn the source voltage ON. (V Output indicator turns
ON.)
Step 6. Record the multimeter reading on the calculation sheet. (Initial output voltage setting is 0
V. )
Step 7. Press [V Output] to turn the source voltage OFF. (V Ouptut indicator turns OFF.)
Step 8. Calculate the test result according to the calculation sheet, and record the result into the
performance test record.
Step 9. Perform this test for all the voltage settings listed in Table 2. The source voltage must be
turned OFF after each test. The source voltage can be changed using the following
procedure :
1. Press [Source Voltage] to display the source voltage setup menu.
2. Use the numeric keys to enter a desired voltage, and press [Enter].
Switch Setting
ID1 0
ID2 0
ID3 0
ID4 1
Output Enable
4339boe1001
ID
1234
Output
4339B
Shunt Connector
RedBlack
Black
Red
V Measurement
Adapter
Multimeter
Banana-Banana Jumper

Other manuals for Agilent Technologies 4339B

Related product manuals