C
Copyright 2008 Agilent Technologies 
Agilent 4339B High Resistance Meter Operation Manual
Manual Change
Agilent Part No. N/A
Sep 2008
Change 1
Change the Measurement Uncertainty value of Source Voltage Accuracy Test (Page 9-19) to the following.
Source
Voltage
Setting
Test Limits Test Result
1
Measurement
Uncertainty
0 V ±0.10 V V ± 0.000206 V
10 V ±0.12 V V ± 0.00027 V
25 V ±0.14 V V ± 0.00054V
50 V ±0.18 V V ± 0.0010 V
100 V ±0.26 V V ± 0.0019 V
200 V ±0.42 V V ± 0.0037 V
201 V ±0.82 V V ± 0.0040 V
250 V ±0.90 V V ± 0.0053 V
500 V ±1.3 V V ± 0.0097 V
1000 V ±2.1 V V ± 0.0188 V
Change 2
Change the Measurement Uncertainty value of Ammeter Offset Voltage Test (Page 9-19) to the following.
Test Limits Test Result
1
Measurement
Uncertainty
±0.5 mV mV ± 0.0083 mV
Change 3
Change the Measurement Uncertainty value of Ammeter Input Resistance Test (Page 9-19) to the
following.
Test Limits Test Result
1
Measurement
Uncertainty
±50 ohm
ohm
± 4.37 ohm