Copyright 2008 Agilent Technologies 
Change 4
Change the Measurement Uncertainty value of Current Measurement Accuracy Test (Page 9-20) to the
following.
Test
Current
Test Limits Measurement
Time
Test Result
1
Measurement
Uncertainty
100 pA ±2.6 pA Long pA ± 0.27 pA
1 nA ±0.0094 nA Long nA ± 0.0007 nA
10 nA ±0.063 nA Long nA ± 0.0074 nA
100 nA ±0.53 nA Long nA ± 0.046 nA
1 A ±0.0043 A Long A ± 0.0005 μA
10 A ±0.036 A Long A ± 0.0053 μA
100 A ±0.60 A Long A ± 0.056 μA
Change 5
Change the Measurement Uncertainty value of Resistance Measurement Accuracy Test (Floating Device)
(Page 9-20) to the following.
Resistor
Setting
Voltage
Setting
Measurement
Time
Test Limits Test Result
1
Measurement
Uncertainty
10
6
ohm 100 V Short ±0.0086 x 10
6
ohm x 10
6
ohm ± 0.00015 × 10
6
ohm
10
7
ohm 100 V Long ±0.0063 x 10
7
ohm x 10
7
ohm ± 0.0003 × 10
7
ohm
10
8
ohm 100 V Long ±0.0073 x 10
8
ohm x 10
8
ohm ± 0.0006 × 10
8
ohm
10
9
ohm 100 V Long ±0.0093 x 10
9
ohm x 10
9
ohm ± 0.0007 × 10
9
ohm
10
10
ohm 100 V Long ±0.0273 x 10
10
ohm x 10
10
ohm ± 0.0012 × 10
10
ohm
10
11
ohm 100 V Long ±0.0453 x 10
11
ohm x 10
11
ohm ± 0.0035 × 10
11
ohm
10
11
ohm 100 V Short ±0.0550 x 10
11
ohm x 10
11
ohm ± 0.0035 × 10
11
ohm
10
11
ohm 10 V Long ±0.0546 x 10
11
ohm x 10
11
ohm ± 0.0035 × 10
11
ohm
Change 6
Change the Measurement Uncertainty value of Resistance Measurement Accuracy Test (Grounded Device)
(Page 9-20) to the following.
Resistor
Setting
Voltage
Setting
Test Limits Test Result
1
Measurement
Uncertainty
10
7
ohm 100 V ±0.0065 x 10
6
ohm x 10
7
ohm ± 0.0003 × 10
7
ohm
10
11
ohm 10 V ±0.0573 x 10
11
ohm x 10
11
ohm ± 0.0035 × 10
11
ohm