4339B
P
erformance
Tests
3. Press
the 3458A
Multimeter's
4
DCV
5
to
set the
measurement mode
to DC
voltage.
4. Set
the V
Measurement A
dapter's ID
switches and
Output
switch
as
follows:
Switch Setting
ID1 0
ID2 0
ID3 0
ID4 1
Output Enable
5.
On the
4339B
,
press
to
turn the
source
voltage
ON.
(V
Output
indicator
turns
ON.)
6.
Record
the
multimeter
reading
on
the
calculation
sheet.
(Initial
output
voltage
setting
is
0
V
.)
7.
Press
to
turn
the
source
voltage
OFF
.
(V
Output
indicator
turns
OFF
.)
8.
Calculate
the
test
result
according
to
the
calculation
sheet,
and
record
the
result
into
the
performance
test
record.
9.
P
erform
this
test
for
all
the
voltage
settings
listed
in T
able 9-2
.
The
source
voltage
must
be
turned
OFF
after
each
test.
The
source
voltage
can
be
changed using
the following
procedure:
a.
Press
to
display
the
source
voltage
setup
menu.
b.
Use
the
numeric
keys
to enter
a
desired
voltage
,
and
press
.
c.
The
desired
voltage
will
be
displayed.
T
able
9-2.
Source
V
oltage
A
ccuracy
T
est
Settings
Source
V
oltage
Setting
0
V
10
V
25
V
50 V
100 V
200 V
201 V
250 V
500 V
1000 V
Maintenance 9-5