EasyManua.ls Logo

Agilent Technologies 4339B - Page 236

Agilent Technologies 4339B
287 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
P
erformance
T
ests 4339B
Ammeter
Oset
V
oltage
and
Input
Resistance
T
est
The
4339B's
ammeter
oset
voltage
and
input
resistance
are
measured.
Specication
Ammeter
Oset
V
oltage:
0.5
mV
Ammeter
Input
Resistance:
1
k
6
5%
T
est
Equipment
Description
Recommended Model
Multimeter
3458A
RC
Box A
dapter
PN
04339-65005
Dual
Banana-BNC(m) Cable
PN
11001-60001
Procedure
Ammeter
Oset
V
oltage
T
est.
1.
Reset the
4339B using
the following
procedure:
a.
Press
to
display the
system reset
menu.
b.
Select
Yes
using
or and
press
.
2.
Set
up the
equipment
as
shown
in
Figure
9-2
.
Figure 9-2. Ammeter Oset Voltage Test Setup
3. Set the RC Box Adapter switch to the
F
position.
4. Press the 3458A Multimeter's
4
DCV
5
to set the measurement mode to DC voltage.
9-6 Maintenance

Other manuals for Agilent Technologies 4339B

Related product manuals