4339B
P
erformance
Tests
5.
Press the
4339B's
to
set the
measurement time
to SHORT
.The
current measurement
time setting
is
indicated
by
the
Meas
Time
annunciator(
9
).
6. Set
the
measurement
range
to
100
A
using
the
following
procedure:
a. Press
to display
the
measurement
range
setup
menu.
b.
Press
until
100
A
appears
,
and
press
.
7.
Press
to
set
the
trigger mode
to Manual.
The current
trigger mode
setting
is
indicated
by
the
Trigger
annunciator(
9
).
8.
Record
the
multimeter
reading
into
the
performance
test
record.
Do
not
change
the
equipment
setup
for
the
following
ammeter
input
resistance
test.
Ammeter
Input
Resistance
T
est.
9.
Record
the
16340A
RC
Box's
10
4
calibration
value
and
the
ammeter
oset
voltage
test
result
on
the
calculation
sheet.
10.
Connect
the
RC
Box
to
the
RC
Box
adapter
as
shown
in
Figure
9-3
.
Figure
9-3.
RC
Box
Connection
11.
Set
the
RC
Box
resistor
to
10
4
.
12.
Set
the
source
voltage
to
1
V
using
the
following
procedure:
a.
Press
to
display
the
source
voltage
setup
menu.
b. Press
to set the
source voltage to 1 V
.
c. Conrm that
+1V
is displayed on the
right side of the LCD
.
13. Press
to turn the source voltage ON. (V Output indicator turns ON.)
14. Record
the multimeter reading on the calculation sheet on the L V
oltage line
.
15. Disconnect the Dual Banana-BNC(m) cable from the Monitor L terminal and connect it to
the Monitor H terminal.
16. Record the multimeter reading on the calculation sheet on the H Voltage line.
17. Calculate the test result according to the calculation sheet, and record the result into the
performance test record.
Maintenance 9-7