P
erformance
T
ests 4339B
Resistance
Measurement
A
ccuracy T
est
The
4339B
measures
the
resistance values
of the
calibrated RC
Box, and
these measurements
are
compared
with
the
RC Box's
calibration values
.
Specication
Basic
Measurement
A
ccuracy:
6
0.6
% (See
Chapter 8
General Information
for
details
.)
Test
Equipment
Description
Recommended
Model
RC
Box
16340A
RC
Box
A
dapter
PN
04339-65005
Procedure
Warning
When
the
High
V
oltage
indicator
is
lit,
the 4339B
outputs
high
voltages
of
up
to
1000
Vdc
maximum.
Resistance
Measurement
A
ccuracy
T
est
(Floating
Device).
1.
Record
the
16340A
RC
Box
calibration
values
into
the
Calculation
Sheet.
2.
Reset
the
4339B using
the
following
procedure:
a.
Press
to
display
the
system
reset menu.
b.
Select
Yes
using
or and
press
.
3.
Set
the
Oset-error
Canceling
to
ON
using
the following
procedure:
a.
Press to
display
the
conguration
setting
menu.
b. Select
Offset
using
or and press .
c.
Select
On
using
or and
press
.
d. Select
Exit
and press to exit the conguration setting menu.
4. Set up the equipment as shown in Figure 9-6.
9-12 Maintenance