4339B
P
erformance
Tests
Figure
9-6.
Resistance
Measurement
A
ccuracy
T
est
Setup
(Floating)
5.
Set
the
RC
Box
A
dapter
switch
to
the
F
position.
6.
Press
.
Select
ExecCal
and
press
to
perform
the
calibration.
7.
Set
the
source
voltage
to
100 V
using
the
following
procedure:
a.
Press
to
display
the
source
voltage
setup
menu.
b.
Press
to
set
the
source
voltage
to
100
V
.
c.
Conrm
that
+
100
V
is
displayed
on
the
right side
of the
LCD.
8.
Press
to
turn
the
source
voltage ON.
(V Output
indicator turns
ON.)
9. Press
. Select
OpenMeas
and
press
to perform
the OPEN
correction.
10. Press
to turn the source voltage OFF
. (V Output indicator turns OFF
.)
11. Connect the RC Box to the RC Box A
dapter as shown in Figure 9-7
Maintenance 9-13