erformance
T
ests 4339B
Figure
9-7.
RC
Box
Connection
(Floating)
12.
Press
to
set
the
measurement
time
to
SHORT
.
The
current
measurement
time
setting
is
indicated
by
the
Meas
Time
annunciator(
9
).
13.
Press
to
set
the
trigger
mode
to
Manual.
The
current
trigger mode
setting
is
indicated by
the
Trigger
annunciator(
9
).
14.
Set
the
RC
Box
resistor
to
10
6
.
15.
Press
to
turn
the
source
voltage
ON.
(V
Output
indicator
lights
.)
16.
Press to
measure
.
17.
Record
the
4339B
reading
on
the
calculation
sheet.
18.
Press
to
turn
the
source
voltage
OFF
.
(V
Output
indicator
turns
OFF
.)
19.
Calculate
the
test
result
according
to
the
calculation
sheet,
and record
the result
into the
performance
test
record.
20.
P
erform
the
test for
all settings
listed in
T
able
9-4
.
The
source
voltage
must
be
turned
OFF
after
each
test.
In
the test
for the
10
11
resistor
,
connect
the RC
Box A
dapter to
the 10
11
terminals
and
set
the
10
11
switch
to
the
BNC
connector position
as shown
in Figure
9-8.
Figure 9-8. RC Box Connection for 10
11
Resistor
9-14 Maintenance