4339B
P
erformance
Tests
T
able 9-4.
Resistance Measurement
A
ccuracy
T
est
Settings
Resistor
Setting
V
oltage
Setting
Measurement
Time
10
6
100 V Short
10
7
100
V
Long
10
8
100
V
Long
10
9
100 V Long
10
10
100
V
Long
10
11
100
V
Long
10
11
100
V
Short
10
11
10
V
Long
Resistance
Measurement
A
ccuracy
T
est
(Grounded
Device).
21. Set
up the
equipment as
shown
in
Figure
9-9
.
Figure 9-9. Resistance Measurement A
ccuracy T
est Setup (Grounded)
22. Set the RC Box A
dapter switch to the
G
position.
23. Set the source voltage to 100
V using the following procedure:
a. Press
to display the source voltage setting.
b. Press to set the source voltage to 100 V.
c. Conrm that
+ 100 V
is displayed on the right side of the LCD.
Maintenance 9-15