24. Press
to turn
the source
voltage ON.
(V Output
indicator turns
ON.)
25.
Press
.
Select
OpenMeas
and press
to
perform the
OPEN correction.
26.
Press
to
turn
the
source
voltage
OFF
.
(V Output
indicator turns
OFF.)
27.
Connect
the
RC
Box
to
the RC
Box A
dapter as
shown in
Figure
9-10
.
Figure 9-10.
RC
Box
Connection
(Grounded)
28.
Set
the
RC
Box
resistor
to
10
7
.
29.
Press
to
turn
the
source
voltage
ON.
(V
Output
indicator
turns
ON.)
30.
Press to
measure
.
31.
Record
the
4339B reading
on the
calculation
sheet.
32.
Press
to
turn
the
source
voltage
OFF
.
(V
Output
indicator
turns
OFF
.)
33.
Calculate
the
test
result
according
to
the
calculation
sheet,
and record
the result
into the
performance
test
record.
34.
P
erform
this
test
for
all
settings listed
in
T
able
9-5
.
The
source
voltage
must
be
turned
OFF
after each test.
Table 9-5.
Resistance
Measurement A
ccuracy Test Settings (Grounded)
Resistor
Setting
Voltage
Setting
10
7
100 V
10
11
10 V
9-16 Maintenance

Other manuals for Agilent Technologies 4339B